MultiEval
SCPI Commands :
CONFigure:WCDMa:MEASurement<instance>:MEValuation:TOUT
CONFigure:WCDMa:MEASurement<instance>:MEValuation:MSCount
CONFigure:WCDMa:MEASurement<instance>:MEValuation:PSLot
CONFigure:WCDMa:MEASurement<instance>:MEValuation:SYNCh
CONFigure:WCDMa:MEASurement<instance>:MEValuation:MOEXception
CONFigure:WCDMa:MEASurement<instance>:MEValuation:SCONdition
CONFigure:WCDMa:MEASurement<instance>:MEValuation:REPetition
- class MultiEvalCls[source]
MultiEval commands group definition. 75 total commands, 11 Subgroups, 7 group commands
- get_mo_exception() bool[source]
# SCPI: CONFigure:WCDMa:MEASurement<instance>:MEValuation:MOEXception value: bool = driver.configure.wcdmaMeas.multiEval.get_mo_exception()
Specifies whether measurement results that the CMP180 identifies as faulty or inaccurate are rejected.
- return:
meas_on_exception: OFF: Faulty results are rejected. ON: Results are never rejected.
- get_ms_count() int[source]
# SCPI: CONFigure:WCDMa:MEASurement<instance>:MEValuation:MSCount value: int = driver.configure.wcdmaMeas.multiEval.get_ms_count()
Selects the total number of measured slots.
- return:
slot_count: No help available
- get_pslot() int[source]
# SCPI: CONFigure:WCDMa:MEASurement<instance>:MEValuation:PSLot value: int = driver.configure.wcdmaMeas.multiEval.get_pslot()
Selects the slot where the CMP180 calculates the results of single slot measurements: ACLR, emission mask, EVM vs chip, CD monitor. The number of the preselected slot must be smaller than the number of measured slots (method RsCMPX_WcdmaMeas. Configure.WcdmaMeas.MultiEval.msCount) .
- return:
slot_number: No help available
- get_repetition() Repeat[source]
# SCPI: CONFigure:WCDMa:MEASurement<instance>:MEValuation:REPetition value: enums.Repeat = driver.configure.wcdmaMeas.multiEval.get_repetition()
Specifies the repetition mode of the measurement. The repetition mode specifies whether the measurement is stopped after a single shot or repeated continuously. Use CONFigure:..:MEAS<i>:…:SCOunt to determine the number of measurement intervals per single shot.
- return:
repetition: SINGleshot: Single-shot measurement CONTinuous: Continuous measurement
- get_scondition() StopCondition[source]
# SCPI: CONFigure:WCDMa:MEASurement<instance>:MEValuation:SCONdition value: enums.StopCondition = driver.configure.wcdmaMeas.multiEval.get_scondition()
Qualifies whether the measurement is stopped after a failed limit check or continued. SLFail means that the measurement is stopped and reaches the RDY state when one of the results exceeds the limits.
- return:
stop_condition: NONE: Continue measurement irrespective of the limit check SLFail: Stop measurement on limit failure
- get_synch() SlotNumber[source]
# SCPI: CONFigure:WCDMa:MEASurement<instance>:MEValuation:SYNCh value: enums.SlotNumber = driver.configure.wcdmaMeas.multiEval.get_synch()
Selects a slot number within the UL WCDMA frames (0 to 14) that the CMP180 displays as the first slot in the measurement interval.
- return:
slot_number: ANY: No frame synchronization SL0 … SL14: First slot = slot 0 … slot 14
- get_timeout() float[source]
# SCPI: CONFigure:WCDMa:MEASurement<instance>:MEValuation:TOUT value: float = driver.configure.wcdmaMeas.multiEval.get_timeout()
Defines a timeout for the measurement. The timer is started when the measurement is initiated via a READ or INIT command. It is not started if the measurement is initiated manually. When the measurement has completed the first measurement cycle (first single shot) , the statistical depth is reached and the timer is reset. If the first measurement cycle has not been completed when the timer expires, the measurement is stopped. The measurement state changes to RDY. The reliability indicator is set to 1, indicating that a measurement timeout occurred. Still running READ, FETCh or CALCulate commands are completed, returning the available results. At least for some results, there are no values at all or the statistical depth has not been reached. A timeout of 0 s corresponds to an infinite measurement timeout.
- return:
timeout: No help available
- set_mo_exception(meas_on_exception: bool) None[source]
# SCPI: CONFigure:WCDMa:MEASurement<instance>:MEValuation:MOEXception driver.configure.wcdmaMeas.multiEval.set_mo_exception(meas_on_exception = False)
Specifies whether measurement results that the CMP180 identifies as faulty or inaccurate are rejected.
- param meas_on_exception:
OFF: Faulty results are rejected. ON: Results are never rejected.
- set_ms_count(slot_count: int) None[source]
# SCPI: CONFigure:WCDMa:MEASurement<instance>:MEValuation:MSCount driver.configure.wcdmaMeas.multiEval.set_ms_count(slot_count = 1)
Selects the total number of measured slots.
- param slot_count:
No help available
- set_pslot(slot_number: int) None[source]
# SCPI: CONFigure:WCDMa:MEASurement<instance>:MEValuation:PSLot driver.configure.wcdmaMeas.multiEval.set_pslot(slot_number = 1)
Selects the slot where the CMP180 calculates the results of single slot measurements: ACLR, emission mask, EVM vs chip, CD monitor. The number of the preselected slot must be smaller than the number of measured slots (method RsCMPX_WcdmaMeas. Configure.WcdmaMeas.MultiEval.msCount) .
- param slot_number:
No help available
- set_repetition(repetition: Repeat) None[source]
# SCPI: CONFigure:WCDMa:MEASurement<instance>:MEValuation:REPetition driver.configure.wcdmaMeas.multiEval.set_repetition(repetition = enums.Repeat.CONTinuous)
Specifies the repetition mode of the measurement. The repetition mode specifies whether the measurement is stopped after a single shot or repeated continuously. Use CONFigure:..:MEAS<i>:…:SCOunt to determine the number of measurement intervals per single shot.
- param repetition:
SINGleshot: Single-shot measurement CONTinuous: Continuous measurement
- set_scondition(stop_condition: StopCondition) None[source]
# SCPI: CONFigure:WCDMa:MEASurement<instance>:MEValuation:SCONdition driver.configure.wcdmaMeas.multiEval.set_scondition(stop_condition = enums.StopCondition.NONE)
Qualifies whether the measurement is stopped after a failed limit check or continued. SLFail means that the measurement is stopped and reaches the RDY state when one of the results exceeds the limits.
- param stop_condition:
NONE: Continue measurement irrespective of the limit check SLFail: Stop measurement on limit failure
- set_synch(slot_number: SlotNumber) None[source]
# SCPI: CONFigure:WCDMa:MEASurement<instance>:MEValuation:SYNCh driver.configure.wcdmaMeas.multiEval.set_synch(slot_number = enums.SlotNumber.ANY)
Selects a slot number within the UL WCDMA frames (0 to 14) that the CMP180 displays as the first slot in the measurement interval.
- param slot_number:
ANY: No frame synchronization SL0 … SL14: First slot = slot 0 … slot 14
- set_timeout(timeout: float) None[source]
# SCPI: CONFigure:WCDMa:MEASurement<instance>:MEValuation:TOUT driver.configure.wcdmaMeas.multiEval.set_timeout(timeout = 1.0)
Defines a timeout for the measurement. The timer is started when the measurement is initiated via a READ or INIT command. It is not started if the measurement is initiated manually. When the measurement has completed the first measurement cycle (first single shot) , the statistical depth is reached and the timer is reset. If the first measurement cycle has not been completed when the timer expires, the measurement is stopped. The measurement state changes to RDY. The reliability indicator is set to 1, indicating that a measurement timeout occurred. Still running READ, FETCh or CALCulate commands are completed, returning the available results. At least for some results, there are no values at all or the statistical depth has not been reached. A timeout of 0 s corresponds to an infinite measurement timeout.
- param timeout:
No help available
Cloning the Group
# Create a clone of the original group, that exists independently
group2 = driver.configure.wcdmaMeas.multiEval.clone()