Current
SCPI Commands :
FETCh:WCDMa:MEASurement<instance>:MEValuation:TRACe:PHD:CURRent
READ:WCDMa:MEASurement<instance>:MEValuation:TRACe:PHD:CURRent
- class CurrentCls[source]
Current commands group definition. 2 total commands, 0 Subgroups, 2 group commands
- fetch() List[float][source]
# SCPI: FETCh:WCDMa:MEASurement<instance>:MEValuation:TRACe:PHD:CURRent value: List[float] = driver.wcdmaMeas.multiEval.trace.phd.current.fetch()
Returns the values of the phase discontinuity traces for up to 120 slots. One value per measured slot is returned (see method RsCMPX_WcdmaMeas.Configure.WcdmaMeas.MultiEval.msCount) .
INTRO_CMD_HELP: The meaning of the value depends on the measurement period (see method RsCMPX_WcdmaMeas.Configure.WcdmaMeas.MultiEval.Mperiod.modulation) :
For full-slot measurements, each value indicates the phase discontinuity at the boundary between a slot and the previous slot. As there is no previous slot for slot 0, the first returned phase discontinuity value equals NCAP.
For half-slot measurements, each value indicates the phase discontinuity at the boundary between the first and second half-slot of a slot. This value can be measured for all slots, including slot 0.
Suppressed linked return values: reliability
- return:
phase_disc: One value per measured slot
- read() List[float][source]
# SCPI: READ:WCDMa:MEASurement<instance>:MEValuation:TRACe:PHD:CURRent value: List[float] = driver.wcdmaMeas.multiEval.trace.phd.current.read()
Returns the values of the phase discontinuity traces for up to 120 slots. One value per measured slot is returned (see method RsCMPX_WcdmaMeas.Configure.WcdmaMeas.MultiEval.msCount) .
INTRO_CMD_HELP: The meaning of the value depends on the measurement period (see method RsCMPX_WcdmaMeas.Configure.WcdmaMeas.MultiEval.Mperiod.modulation) :
For full-slot measurements, each value indicates the phase discontinuity at the boundary between a slot and the previous slot. As there is no previous slot for slot 0, the first returned phase discontinuity value equals NCAP.
For half-slot measurements, each value indicates the phase discontinuity at the boundary between the first and second half-slot of a slot. This value can be measured for all slots, including slot 0.
Suppressed linked return values: reliability
- return:
phase_disc: One value per measured slot